Medical equipment maintenance in a GP practice: keeping a register

Practice Managed team5 min read

Every practice has a defibrillator, scales, blood pressure monitors and other equipment that must work when it is needed. An equipment register is one list that shows each item, when it was last serviced or calibrated, and when it is next due. This guide covers what the Standards ask for, what to record, and the items that need extra care.

What the Standards ask for

5th edition (what your next visit uses): criterion GP5.2, Practice equipment, asks that the practice maintains its clinical equipment in line with each maker's recommendations, and has a defibrillator. The guidance says equipment is serviced regularly. It suggests a register or a maintenance log with receipts, and recording the dates of servicing.

6th edition (what is coming): criterion CG11, Practice equipment, goes further. The practice has all the required equipment on its list, with each item maintained and calibrated annually. The list includes a defibrillator, an electrocardiograph, blood pressure monitors, scales, a thermometer, a pulse oximeter, oxygen, emergency medicines, a height-adjustable bed and a doctor's bag. CG11 also asks that:

  • the clinical team can use the equipment safely and effectively
  • the practice has timely access to a spirometer

The guidance says equipment that needs calibration, has consumables with expiry dates, or runs on power or batteries is serviced every year, following the maker's instructions. It suggests a central register or a checklist in each consulting room, and keeping the paperwork from the companies that test and calibrate your equipment.

What to record for each item

From the RACGP's guidance, a register needs:

  • The item, its make and model, and its serial or asset number
  • Where it is kept and who looks after it
  • How often the maker says to service it
  • The date it was last serviced or calibrated, who did it and the result
  • Who approved it for use again, after a repair or service
  • When it is next due
  • Whether the certificate or receipt is kept

Start by walking through every room and listing what is there. Delete the items you don't have and add what you do. Then fill in the dates from the paperwork you already hold. Anything with no record is the first job.

Items that need extra care

The defibrillator

Check its pads and battery as well as the device, because both expire. The Standards don't set a schedule for checks beyond following the maker's instructions, so use the maker's.

The steriliser

If you sterilise instruments, the RACGP's infection control guidelines ask for more:

  • Validation at least once a year, and after installation, moving it, a major repair or a big change to the process. A qualified technician does it and gives you a detailed report.
  • Maintenance records that show the reason, the date, the model and serial number, and who approved it for use again. Keep them where you can find them.
  • Don't use it until all required maintenance is done and recorded.

The guidelines note that the Australian Standard for office-based practices, AS 5369, allows maintenance up to 10 working days after the scheduled date, and that your practice needs a policy on how much delay it accepts.

The vaccine fridge and its data logger

Victoria's Department of Health advises calibrating thermometers, data loggers and sensors and changing their batteries every year. For the daily checks, see our vaccine fridge checklist.

The doctor's bag

Its contents and expiry dates need regular checks too. See our doctor's bag guide.

Electrical test and tag

Testing and tagging under the Australian Standard AS/NZS 3760 is for office plug-in equipment such as kettles, heaters and computers. It does not cover medical devices or electrical equipment in patient care areas: for those, follow each maker's instructions. How often office equipment must be tested depends on where it is used, so ask a competent person or your state's work safety regulator.

Training the team

CG11 asks that the clinical team can use the equipment safely. Keep a record of who has been shown how to use each item, for example in your staff training register, and include it in induction.

A simple routine

  • When something is serviced: update its row the same day and file the certificate.
  • Once a month: check what is due in the next 60 days and book it.
  • Once a year: walk the practice again and check the list is complete.

Our free equipment maintenance register has the RACGP's equipment already listed, a column for who approved each item for use again, a Status column that shows what is overdue or due soon, and a second tab for electrical test and tag.

We check every fact against official sources, but rules change. Always check the official source and each maker's instructions before you rely on it.

Want to know where the rest of your practice stands? Take the free readiness check: ten yes-or-no questions, your score straight away, and a to-do list sent to your inbox.