Equipment maintenance register template

One row per item of equipment. Type the next due date and the register shows what is overdue or due soon.

The first page of the Equipment maintenance register template
The first page of the template, as you'll download it. Click the picture to see it full size.

What's inside

  • Register: item, make, serial number, location, who is responsible, how often the maker says to service it, last serviced, result, who approved it for use again, next due and status.
  • Starts with equipment the RACGP Standards name, such as the defibrillator, ECG, spirometer and vaccine fridge.
  • Electrical test and tag: a separate tab for office plug-in equipment.

Who fills it in, and when

Each item has a person responsible, who updates its row whenever it is serviced or calibrated. The practice manager checks the Status column once a month. Use the test and tag tab each time office plug-in equipment is tested. Start by deleting the rows for equipment you don't have and adding what you do.

How to use it

  1. Delete the items you don't have and add the ones you do.
  2. Fill in when each was last serviced and when it's next due.
  3. Check the Status column once a month and keep the service paperwork.

Equipment maintenance register

Excel · 48 KB · free

Which RACGP Standards it helps with

Supports 5th edition criterion GP5.2 (Practice equipment) and 6th edition criterion CG11 (Practice equipment). The file notes the 6th edition asks for each item to be maintained and calibrated every year. The 6th edition also asks that the clinical team can use the equipment safely, so keep a record of their training too.

An example

Example only: Spirometer, room 3, person responsible A. Smith (nurse). Last calibrated 2 February 2026 by the supplier, result passed, next due 2 February 2027, certificate kept: Yes. When February 2027 is within 60 days, the Status column changes to Due soon. On the test and tag tab, a row might read: kettle, staff room, tested by an electrician on 5 May 2026, passed.

Common mistakes to avoid

  • Using the test and tag tab for medical devices: follow each maker's instructions for those.
  • Throwing away the paperwork from the companies that test and calibrate your equipment.
  • Missing items with batteries, pads or parts that expire, such as the defibrillator pads and battery.

Goes well with

It goes with our guide Medical equipment maintenance in a GP practice: keeping a register.

Want to know where the rest of your practice stands? Take the free readiness check: ten yes-or-no questions, your score straight away, and a to-do list sent to your inbox.